Parameters | Proposed Methods | Reported method |
---|
HPLC | HPTLC | HPLC(19)* | HPTLC(18)** |
---|
Mean Recovery | 100.02%±0.39 | 99.50%±0.74 | 99.81%±0.47 | 99.11%±0.80 |
n | 9 | 9 | 9 | 9 |
SD | 0.39 | 0.74 | 0.47 | 0.80 |
Variance | 0.156 | 0.553 | 0.225 | 0.654 |
Student t-test | 1.08(2.12) | 1.09(2.12) | | |
F-Test | 1.44(2.44) | 1.18(2.44) | | |
- * RP-HPLC using C18, end-capped (250 × 4.6) mm, 5 μm, at 250 nm. The mobile phase A was buffer pH 3.0; while, the mobile phase B was acetonitrile (ACN) (100%) in gradient mode flow rate 1 ml/min